YS-Y127 Dentin Barrier Cytotoxicity Test Kit

YS-Y127 Dentin Barrier Cytotoxicity Test Kit

【Scope of application】
Suitable for oral materials that directly contact dentin, such as dental resins, adhesives, filling materials, and base materials; biological safety verification during material R&D stages; quality inspection for medical device registration (e.g., Class III dental materials); simulating the impact of dentin diffusion on pulp cells. 【Relevant Standards】
    Complies with YY/T 0127.18-2016
【Technical Parameters】 1.Temperature Control Range of Heating Stage: 30-40℃ for chambers; 2.Flow Rate Range of Peristaltic Pump: Not less than 0.005-500mL/min; 3.Volume of Culture Medium Bottle and Waste Liquid Bottle: Not less than 100mL each; 4.Material of Dentin Slice Holder and Cell Culture Perfusion Chamber: 316 stainless steel; 5.Lumen Dimensions of Perfusion Chamber: Diameter ≥6mm, Height ≥20mm. 【Accessories List】 Cell Culture Inserts: 12mm, Polycarbonate (PC), 3.0 µm, 1 pack.