【Scope of application】
The water resistance measuring instrument is mainly used for the determination of the water resistance of contact wound dressings (500mm hydrostatic pressure 300S capacity). 【Relevant Standards】
    YY/T 0471.3-2004
【Scope of application】: The water resistance measuring instrument is mainly used for the determination of the water resistance of contact wound dressings (500mm hydrostatic pressure 300S capacity). [Meet the standard]: YY/T 0471.3-2004 【Technical Parameters】: 1. Liquid level scale: 500mm 2. Inner diameter of the fixture: 50mm 3. Fixture depth: 40mm 4. Fixture outer diameter: 80mm 5. Water pipe inner diameter: 10mm 6. Water pipe height: 550mm 【Test Steps】: 1. The prepared sample should be free of wrinkles; 2. Fill the pool with purified water at (21±2)°C; 3. Place the sample on the lower ring in a horizontal sliding manner to avoid air between the surface of water and the lower surface of the sample; 4. Cover the upper surface of the sample with dry filter paper larger than the test area, put the upper ring, and clamp it with a threaded device; 5. Inject water into the tube until it reaches the required water level above the sample surface; 6. Maintain the water pressure (300±10)S; 7. Check whether there is water seepage through the sample on the filter paper and record the results; 【Evaluation of test results】: Check whether the filter paper has passed the water seepage and record the results, if any of the 3 samples have water seepage test failed.