【Scope of application】
Experiments for testing liquid infiltration and loss of nonwovens
【Relevant Standards】
GB/T242185.11、ISO 9073.11等
【Scope of application】:
Experiments for testing liquid infiltration and loss of nonwovens
[Meet the standard]:
GB/T242185.11、ISO 9073.11等
【Technical Parameters】:
1. There are 2 black reference wires marked on the test bench, and the spacing is 250±0.2mm. The low line is from the end of the test bench to 3±0.2mm.
The position of the absorption medium at the end; The high line is the position of the center line of the draining test tube, about 25mm down from the top of the test specimen;
2. The inclination angle of the test bench: 25 degrees;
3. There is a clip symmetrical reference mark at the test bench (40±0.2) mm;
4. The center position of the equipment can ensure the axial release of liquid in the test tube;
5. The support frame with standard absorbent liner is arranged at the low end of the test specimen;
6. Drip port inner diameter: 5mm;
7. With a drip and loss length of 250±0.2mm, the drip device can be transmitted in a continuous liquid state within a time of (4±0.1)s
(25±0.5) g of test liquid through a glass test tube;
8. Overall dimensions: 260×510×450mm;
9. Net weight: 5kg.