【Scope of application】
Suitable for evaluating the antibacterial rate, Minimum Inhibitory Concentration (MIC), and long-acting antimicrobial efficacy of wound dressings containing antibacterial components such as silver ions and chitosan (e.g., hydrogel, hydrocolloid, and foam dressings) against standard bacterial strains including Staphylococcus aureus and Escherichia coli.
【Relevant Standards】
YY/T 1477.1-2016 “Standard Test Models for Evaluating the Performance of Contact Wound Dressings – Part 1: In Vitro Wound Model for Evaluating Antimicrobial Activity”
【Technical Parameters】
1. Peristaltic Pump
1.1 Number of Channels: At least 6 channels;
1.2 Flow Rate: Controllable flow rate of 5ml/h per channel;
1.3 Special Pump Tubing: Capable of withstanding high-temperature sterilization at 120℃ minimum;
2. Reactor
2.1 Reactor: At least 6 channels;
2.2 Weights: 33±2g, 6 sets in total;
【Accessories List】
1.Reagent Bottles: 2 units (each bottle cap is equipped with an interface suitable for connecting pipelines and an interface for installing bacterial filters);
2.Glass Burette: 1 unit;
3.Peristaltic Pump (6-channel): 1 unit;
4.Reactor (6-channel): 1 unit;
5.Special Pump Tubing: 1 piece (length: 10m, pressure steam sterilizable);
6.Three-Way Adapters for Pump Tubing: 10 pieces (pressure steam sterilizable);
7.Glass Burette Fixing Bracket: 1 unit;